Technical
Specifications

Artsens Plus Device

ARTSENS Plus

Artsens Pen Device

ARTSENS Pen

Measurements Supported

End-diastolic diameter
Distension waveform
Local vascular stiffness (β, AC, Ep, PWVβ)
Regional vascular stiffness (Carotid-Femoral PWV)
Central blood pressure
Intigrated brachial blood pressure module
Resolution of distension

10 um

10 um

Resolution of end-diastolic diameter

< 300 um

< 300 um

Systolic blood pressure range

40 - 260 mmHg

-

Diastolic blood pressure range

20 - 200 mmHg

-

Acccuracy of blood pressure

+/- 5 mmHg

-

Software Features

Real-time ultrasound signal analysis at 100 Hz
Ultrasound signal analysis at 500 Hz
Automated diameter cycle extraction
Automated diameter and pressure cycle extraction
Processed data export in xml / excel format for all readings
Report generation & exporting
Over the Air (OTA) software updates
Raw data saving in host PC (Windows)
Beat to beat capture of distension waveforms

Hardware Features

Ultrasound probe

5MHz / 10 MHz, broad band, focused probe

Pulse repetition frequency

upto 500 Hz

upto 100 Hz

Power supply

Internal Battery power

Bus-powered via USB 2.0

Data communication interface

USB 2.0

USB 2.0

Dimensions (hardware module)

240 mm x 170 mm x 60 mm

180 mm x 30 mm x 20 mm

Weight (electronics hardware module)

~ 900 g

~ 55 g

Host Tablet/Laptop /PC Minimum Requirements

Intel i5 Processor, 4GB RAM, 128 GB ROM, Windows 10 OS, 1xUSB (type A) 2.0 port or higher

Intel i3 Processor, 4GB RAM, 64 GB ROM, Windows 10 OS, 1xUSB (type A) 2.0 port or higher

Environmental Conditions

Operating Temperature

0°C to 50°C

0°C to 50°C

Storage Temperature (Commercial grade)

0°C to 70°C

0°C to 70°C

Relative Humidity, non-condensing

10 to 80%

10 to 80%

Regulatory Compliance

IEC 60601-1 (Essential performance & safety)
IEC 60601-1-2 (EMI/EMC)